Among the several available X-ray optics for synchrotron radiation producing micrometre and submicrometre beams with high intensity, the X-ray waveguide (WG) can provide the smallest hard X-ray beam in one direction. A drawback of this optics is that, owing to the divergence at the exit, a nanometre-sized spot on the sample can only be obtained if this is within a few micrometres of the WG exit. Another limitation is that in planar WGs the beam is compressed in only one direction. Here, using a dynamically bent elliptical Si/Pt mirror, the guided X-ray beam has been refocused at approximately 1 m from the waveguide exit. The large working distance between the device and the submicrometre focus leaves some space for sample environment (vacuum chamber, furnace, cryostat, magnets, high-pressure device etc.) and allows cross-coupled geometries with two WGs for efficient compression in two directions.

Fonte
Journal of Synchrotron Radiation
Volume
13
Fascicolo
1
Pagine
85-87
Citazioni
2
ID archivio
8340e9f6213614c41f4de6268d73d626
Riferimenti
Volume: 13 Issue: 1 Pages: 85-87