Design, fabrication and characterization of a capacitive micromachined ultrasonic probe for medical imaging

Data di pubblicazione: 27 Dic 2005

RivistaFonte dati: ARCHIVIO

In this paper we report the design, fabrication process, and characterization of a 64-elements capacitive micromachined ultrasonic transducer (cMUT), 3 MHz center frequency, 100% fractional bandwidth. Using this transducer, we developed a linear probe for application in medical echographic imaging. The probe was fully characterized and tested with a commercial echographic scanner to obtain first images from phantoms and in vivo human body. The results, which quickly follow similar results obtained by other researchers, clearly show the great potentiality of this new emerging technology. The cMUT probe works better than the standard piezoelectric probe as far as the axial resolution is concerned, but it suffers from low sensitivity. At present this can be a limit, especially for in depth operation. But we are strongly confident that significant improvements can be obtained in the very near future to overcome this …

Editore
IEEE
Fonte
IEEE transactions on ultrasonics, ferroelectrics, and frequency control
ID archivio
16930c342b2df449848265712d885007
Riferimenti
Volume: 52 Issue: 12 Pages: 2259-2269