Spectroscopic response of a CdZnTe multiple electrode detector

Data di pubblicazione: 21 Dic 2007

RivistaFonte dati: ARCHIVIO
Autori: Adriano Cola

The spectroscopic performances of a CdZnTe detector (crystal size: 5×5×0.9 mm3) with five electrodes (cathode, anode and three steering electrodes) were studied. The anode layout, which consists of a circular electrode (φ=80 μm) surrounded by two ring electrodes (gap=100 μm; radial width Δr=100 μm) and by one electrode that extends to the edge of the crystal, is mostly sensitive to the electron carriers, overcoming the well known effect of the hole trapping in the measured spectra. We report on the spectroscopic response of the detector at different bias voltages of the electrodes and at various photon energies (109Cd, 241Am and 57Co sources). The CdZnTe detector exhibits excellent energy resolution (0.87 keV FWHM at 22.1 keV, 1.15 keV FWHM at 59.5 keV and 1.91 keV FWHM at 122.1 keV; working temperature −10 °C) and low tailing (FW.1M to FWHM ratio of 1.87 at 22.1 keV, 1.93 at 59.5 keV and 2.01 at …

Editore
North-Holland
Fonte
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
ID archivio
6b07b1dee74426efc0b85b92ec0dcf1f
Riferimenti
Volume: 583 Issue: 2-3 Pages: 324-331