New label free CA125 detection based on gold nanostructured screen-printed electrode

Data di pubblicazione: 31 Mar 2013

RivistaFonte dati: ARCHIVIO
Autori: A Ravalli, G Pilon Dos Santos, Matteo Ferroni, G Faglia, Hideko Yamanaka, G Marrazza

This paper describes the optimisation and the analytical performances of a label-free impedimetric immunosensor for the detection of tumour marker CA125 based on gold nanoparticles modified screen-printed graphite electrode. Experimental conditions of each step for the developed immunosensor were studied and optimised. The immunosensor response varied linearly (r2 = 0.996) with antigen concentration between 0 and 100 U/mL. The estimated detection limit was 6.7 U/mL. The electrochemical immunosensor allowed unambiguous identification of CA125, while no significant non-specific signal was detected in the case of all negative controls. The analytical usefulness of the impedimetric immunosensor was finally demonstrated analysing serum samples.

Editore
Elsevier
Fonte
Sensors and Actuators B: Chemical
ID archivio
46f2f80fab032025c935c980bb6f336b
Riferimenti
Volume: 179 Pages: 194-200