Innovative Chemical Strategy for PCR-free Genetic Detection of Pathogens by an Integrated Electrochemical Biosensor

Data di pubblicazione: 1 Gen 2017

RivistaFonte dati: ARCHIVIO
Autori: Emanuele Luigi LuigiáSciuto, Maria LetiziaáDi Pietro, Maria GraziaáGrimaldi

An innovative chemical strategy integrated in a miniaturized electrochemical device was developed for sensitive detection of pathogen genome (HBV virus) without any amplification step. Results show a Limit of Detection comparable to the standard qRT-PCR method (20 copies/reaction), paving the way to future development of genetic PoC devices addressing automatized and low-cost molecular diagnostics.

Editore
Royal Society of Chemistry
Fonte
Analyst
ID archivio
32a81e300ae5f89e806bbeb18c9a7fcf