Note: Dead time causes and correction method for single photon avalanche diode devices

Data di pubblicazione: 26 Ago 2010

RivistaFonte dati: ARCHIVIO

Single photon avalanche diode (SPAD) is the new generation of Geiger–Muller (GM) detectors, developed with semiconductor technology, and able to detect single photons, mainly in visible range. In this work we study the signal generation process and the dead time (DT) mechanisms of the device under a constant light regime. According to our results, it is possible to discriminate low rate signals from afterpulse and noise production and, moreover, to overcome the saturation effect due to the dead time losses. Starting from hybrid DT model [S. H. Lee and R. P. Gardner, Appl. Radiat. Isot. 53, 731 (2000)] we have been able to evaluate the real amount of incident photon rate up to 107 cps using a passive quenched device with 0.97 μs total dead time. In this way the passive quenched SPAD achieves the same performance of the active quenched one showing that relatively complex data analysis and complex …

Editore
American Institute of Physics
Fonte
Review of Scientific Instruments
ID archivio
afb447800f1f99574ffd168a612a687a
Riferimenti
Volume: 81 Issue: 8 Pages: 086102