Neutron Detection Study through Simulations with Fluka

Data di pubblicazione: 31 Mag 2022

ConvegnoFonte dati: OPENALEXTipo OpenAlex: articleOpen Access
Autori: A. Muoio, Alessandro Meli, Antonio Trotta, Miriam Parisi, Laura Meda, Francesco La Via

In this work the purpose of the simulations is to optimize a new large volume Silicon Carbide (SiC) detector for 14.1 MeV neutrons. The device has an active thickness obtained by epitaxial growth and an active area of 25 mm2. In the first step of the simulations we compare SiC detector performance to Diamond and Silicon detectors, with the same geometric features. In the second step of the simulations we have found the best solution to improve the response of the detector for a fixed epitaxial layer thickness using an overlayer of aniline (C6H7N).

Editore
Trans Tech Publications Ltd
Fonte
Materials Science Forum
Volume
1062
Pagine
509-513
Citazioni
2
ID archivio
0c934f4bdc6bdff2d5c813a9a3488c6b
Riferimenti
Volume: 1062 Pages: 509-513