-A A +A
Type: 
Journal
Description: 
In this work, we present the analytical modeling and preliminary experimental results for the choice of the optimal frequencies when performing amplitude and phase measurements with a scanning microwave microscope. In particular, the analysis is related to the reflection mode operation of the instrument, ie, the acquisition of the complex reflection coefficient data, usually referred as S 11. The studied configuration is composed of an atomic force microscope with a microwave matched nanometric cantilever probe tip, connected by a λ/2 coaxial cable resonator to a vector network analyzer. The set-up is provided by Keysight Technologies. As a peculiar result, the optimal frequencies, where the maximum sensitivity is achieved, are different for the amplitude and for the phase signals. The analysis is focused on measurements of dielectric samples, like semiconductor devices, textile pieces, and biological specimens …
Publisher: 
AIP Publishing
Publication date: 
20 Jul 2015
Authors: 

GM Sardi, A Lucibello, M Kasper, G Gramse, E Proietti, F Kienberger, R Marcelli

Biblio References: 
Volume: 107 Issue: 3
Origin: 
Applied Physics Letters